TYSHAK II ® Percutaneous Transluminal Valvuloplasty catheter
Recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.
- A patient with isolated pulmonary stenosis.
- A patient with valvular pulmonary stenosis with other minor congenital heartdisease that does not require surgical intervention
The TYSHAK II® catheter is a coaxially constructed catheter with a distally mounted non-compliant balloon. The catheter exhibits an extremely low profile that allows the interventionalist to use the small introducer possible. Due to the coaxial construction, the catheter inflates and deflates extremely fast. The TYSHAK II® catheter provides smaller introducers and in some cases slightly lower burst pressures than the TYSHAK®.
- General
- Specifications
-
Catheter Characteristics
The TYSHAK II® catheter is a coaxially constructed catheter with a distally mounted non-compliant balloon. The catheter exhibits an extremely low profile that allows the interventionalist to use the small introducer possible. Due to the coaxial construction, the catheter inflates and deflates extremely fast. The TYSHAK II® catheter provides smaller introducers and in some cases slightly lower burst pressures than the TYSHAK®.
Super Thin Balloon
The TYSHAK II® PTV balloon is super thin for a low deflated profile that maintains tip flexibility. The exceptionally low profile balloon requires the smallest introducer possible. Nominal dimensions are maintained over the entire length of the balloon.
Maximum Trackability
The distal shaft through the balloon is highly flexible for exceptional maneuverability. This combined with the pushability of the coaxial shaft provides outstanding tracking performance.
Radiopaque marker
Platinum marker bands facilitate reliable positioning of the balloon.
-
Product specifications
Balloon
Diameter
(MM)Balloon
Length
(CM)Introducer
Size
(FR)Shaft
Size
(FR)Usable
Length
(CM)Guide
Wire
(Inches)Rated
Burst
(ATM)*Nominal
Pressure
(ATM)Catalog
No.4.0 2.0 4 4 70 0.021 6 4.5 PDC500 4.0 2.0 4 4 100 0.021 6 4.5 SN003 4.0 10.0 4 4 70 0.021 6 4.5 PDC551 5.0 2.0 4 4 70 0.021 6 4.5 PDC501 5.0 2.0 4 4 100 0.021 6 4.5 SN004 5.0 3.0 4 4 70 0.021 6 4.5 SO020 6.0 2.0 4 4 70 0.021 4 3.5 PDC502 6.0 2.0 4 4 100 0.021 4 3.5 SN005 6.0 3.0 4 4 70 0.021 4 3.5 SO021 7.0 2.0 4 4 70 0.021 4 3.5 PDC503 7.0 2.0 4 4 100 0.021 4 3.5 SN006 7.0 3.0 4 4 70 0.021 4 3.5 PDC509 7.0 3.0 4 4 100 0.021 4 3.5 SN010 8.0 2.0 4 4 70 0.021 4 3.5 PDC504 8.0 2.0 4 4 100 0.021 4 3.5 SN007 8.0 3.0 4 4 70 0.021 4 3.5 PDC510 8.0 3.0 4 4 100 0.021 4 3.5 SN011 8.0 4.0 4 4 70 0.021 4 3.5 SO029 8.0 5.0 4 4 70 0.021 4 3.5 SO032 8.0 6.0 4 4 70 0.021 4 3.5 SO030 9.0 2.0 5 5 90 0.025 3.5 3 SO007 9.0 3.0 5 5 90 0.025 3.5 3 PDC505 9.0 3.0 5 5 100 0.025 3.5 3 SN008 10.0 2.0 5 5 90 0.025 3.5 3 SO003 10.0 3.0 5 5 90 0.025 3.5 3 PDC506 10.0 3.0 5 5 100 0.025 3.5 3 SN009 10.0 4.0 5 5 90 0.025 3.5 3 SO001 10.0 5.0 5 5 90 0.025 3.5 3 SO031 10.0 6.0 6 6 100 0.025 3.5 3 SO009 11.0 2.0 5 5 90 0.025 3.5 3 SN11290 11.0 3.0 5 5 90 0.025 3.5 3 SO017 11.0 4.0 5 5 90 0.025 3.5 3 SO002 12.0 2.0 5 5 90 0.025 3.5 3 SO008 12.0 2.0 6 6 90 0.035 3.5 3 SO014 12.0 3.0 5 5 90 0.025 3.5 3 PDC507 12.0 4.0 6 6 90 0.035 3.5 3 PDC508 12.0 5.0 5 5 90 0.025 3.5 3 SO033 12.0 6.0 6 6 90 0.035 3.5 3 SO004 13.0 2.0 6 6 100 0.035 3 2.5 SO035 13.0 3.0 6 6 100 0.035 3 2.5 SO036 13.0 4.0 6 6 100 0.035 3 2.5 SO034 13.0 5.0 6 6 100 0.035 3 2.5 SO069 14.0 2.0 7 7 100 0.035 3 2 SO018 14.0 3.0 7 7 100 0.035 3 2 PDC511 14.0 4.0 7 7 100 0.035 3 2 PDC512 14.0 5.0 7 7 100 0.035 3 2 PDC513 14.0 6.0 7 7 100 0.035 3 2 PDC514 15.0 3.0 7 7 100 0.035 3 2 PDC515 15.0 4.0 7 7 100 0.035 3 2 PDC516 15.0 5.0 7 7 100 0.035 3 2 PDC517 15.0 6.0 7 7 100 0.035 3 2 PDC518 16.0 2.0 7 7 100 0.035 2.5 2 SN002 16.0 3.0 7 7 100 0.035 2.5 2 PDC519 16.0 4.0 7 7 100 0.035 2.5 2 PDC520 16.0 5.0 7 7 100 0.035 2.5 2 PDC521 16.0 6.0 7 7 100 0.035 2.5 2 PDC522 17.0 3.0 7 7 100 0.035 2.5 2 PDC523 17.0 4.0 7 7 100 0.035 2.5 2 PDC524 17.0 5.0 7 7 100 0.035 2.5 2 PDC525 17.0 6.0 7 7 100 0.035 2.5 2 PDC526 18.0 3.0 8 8 100 0.035 2 1.5 PDC527 18.0 4.0 8 8 100 0.035 2 1.5 PDC528 18.0 5.0 8 8 100 0.035 2 1.5 PDC529 18.0 6.0 8 8 100 0.035 2 1.5 PDC530 20.0 2.0 8 8 100 0.035 2 1.5 SN001 20.0 3.0 8 8 100 0.035 2 1.5 PDC531 20.0 4.0 8 8 100 0.035 2 1.5 PDC532 20.0 5.0 8 8 100 0.035 2 1.5 PDC533 20.0 6.0 8 8 100 0.035 2 1.5 PDC534 22.0 2.0 8 8 100 0.035 2 1.5 SN013 22.0 3.0 8 8 100 0.035 2 1.5 PDC535 22.0 4.0 8 8 100 0.035 2 1.5 PDC536 22.0 5.0 8 8 100 0.035 2 1.5 PDC537 22.0 6.0 8 8 100 0.035 2 1.5 PDC538 23.0 3.0 9 9 100 0.035 2 1.5 PDC539 23.0 4.0 9 9 100 0.035 2 1.5 PDC540 23.0 5.0 9 9 100 0.035 2 1.5 PDC541 23.0 6.0 9 9 100 0.035 2 1.5 PDC542 25.0 3.0 9 9 100 0.035 1.5 1 PDC543 25.0 4.0 9 9 100 0.035 1.5 1 PDC544 25.0 5.0 9 9 100 0.035 1.5 1 PDC545 25.0 6.0 9 9 100 0.035 1.5 1 PDC546 30.0 3.0 10 9 100 0.035 1.5 1 PDC547 30.0 4.0 10 9 100 0.035 1.5 1 PDC548 30.0 5.0 10 9 100 0.035 1.5 1 PDC549 30.0 6.0 10 9 100 0.035 1.5 1 PDC550 30.0 10.0 10 9 100 0.035 1.5 1 PDC552 The devices listed have CE mark approval. Any customer or distributor outside of the EU will be subject to their individual country's regulations in regards to the importation and sale of these products.
This publication could include technical or other inaccuracies or typographical errors. Changes are periodically added to the information herein; these changes will be incorporated in the new editions of the publication. NuMED Incorporated may make improvements and/or changes in the product(s) and/or the program(s) described in this publication at any time.